Automatic Potentiometric Titrator LPTT-A10 is an automatic titrator with integrated burette valve design. This modular titrator consist of three units: volume titration device, control device, measurement device. The titration analysis includes acid-base titration, oxidation-reduction titration, non-aqueous, precipitation titration and others.
| Titration mode |
Macro titration, Micro titration, Volume setting titrations, End point titration, Stored titration |
| Analysis method |
Acid-base Titration, Oxidation-reduction titration, Precipitation titration, Complexometric, Non-aqueous titration |
| Measurement Range |
| pH value | 0.00 to 14.00 | | mV value | -1999.0 to 1999.0 mV | | Temperature value | 0 to 100 °C | |
| Resolution |
| pH value | 0.04 pH | | mV value | 0.1 mV | | Temperature value | 0.1°C | |
| Accuracy |
| pH value | ± 0.02 pH | | mV value | ± 0.2 mV | | Temperature value | 0.2°C | |
| Burette Volume Accuracy |
| 15 ml burette | ± 0.025 ml | | 25 ml burette | ± 0.035 ml | | Titration analysis parallelism | 0.002 | | Millivolt accuracy | ≤ 0.2 mV | |
| Stir |
Coil realize Magnetism stir |
| Titration cup |
100 ml |
| Electrodes port type |
mV, pH 1, reference electrode 1, PT1000 electrode 1 |
| Display |
7 inch TFT touch screen display |
| Communication port |
USB, Stir port, power port, printer port |
| Learn titration |
Methods can be edited and stored |
| Power supply |
110 to 240 V, 50/60 Hz |
| External dimension |
530 × 455 × 280 mm |
| Net Weight |
10 kg |
Used in performing iodine peroxide titration and other titration methods in food industry. Also it finds application in pharma, chemical, petrochemical, life science industry.